ChromX®

01) Paracetamol IP Assay
Column: ChromX® C18 5µm, 250x4.6mm, P/N:CHI18502S045-0
Mobile Phase: 85%phosphate Buffer pH 4.0/ 7% ACN
Flow Rate: 1.0 ml/min
Column Oven: 40°C
UV: 215nm
Sample: 0.3mg/mL Paracetamol Standard

Chromatogram showing sharp peak at retention time ~5.1 minutes
Excellent peak shape and resolution

02) Favipiravir IP Assay
Column: ChromX® C18 5µm, 250x4.6mm, P/N:CHI18502S045-0
Mobile Phase: 77%0.1% OPA/53% ACN
Flow Rate: 1.0 ml/min
Column Oven: 25°C
UV: 225nm
Sample: 5mg/100mL Favipiravir Standard

Chromatogram showing well-defined peak at retention time ~4.8 minutes
High sensitivity and selectivity

03) Cefixime Assay USP
Column: ChromX® C18 (2) 4µm, 125x4.6mm, P/N: CHI18402S45-2
Mobile Phase: 25 volume of ACN and 75 volume of tetrabutylammonium hydroxide solution prepared by diluting 25ml of 0.4M tetrabutylammoniumhydroxide solution to 1000 ml with water and adjusted pH to 6.6 with OPA
Flow Rate: 1.4 ml/min
Column Oven: 40°C
UV: 254nm
Injection Volume: 10µL
Sample: Cefixime Reference Solution

Chromatogram showing two distinct peaks (Imp-E and Cefixime)
Excellent separation and baseline resolution

Analytical Results Summary
ID Name Ret. Time Area Height Area% Theoretical Plate# Tailing Factor Resolution RRT
1 Imp-E 9.100 2768481 165678 25.282 6708 1.27 0.00 1.0
2 Cefixime 10.172 8181977 385020 74.718 4330 1.73 2.02 1.0
10950405 630598
04) Ranitidine hydrochloride RS IP
Column: ChromX® C18-AQ 3.5µm, 100X4.6mm
Mobile Phase: A- 2% ACN / 98% buffer (0.05M KH2PO4 pH 7.1) B- 22% ACN / 78% buffer (0.05M KH2PO4 pH 7.1)
Flow Rate: 1.5 ml/min
Gradient:
Time %A %B
0 100 0
10 0 100
15 0 100
16 0 100
20 100 0
Column Temp: Ambient
UV: 230nm
Injection Volume: 10µL
Sample: Reference solution b (Ranitidine HCl + impurity J)

Chromatogram showing two distinct peaks (Imp-E and Cefixime)
Excellent separation and baseline resolution

05) Analysis of Clrotimazol
Column: ChromX® C8 5µm, 150X4.6mm, P/N:CH085015046-0
Mobile Phase: A-1 gm K2HPO4 + 0.5g of tetrabutyammonium hydrogen in 1000ml water, B- ACN
Flow Rate: 1.0 ml/min
Gradient:
Time %A %B
0 75 25
3 75 25
25 20 80
30 20 80
31 75 25
35 75 25
Column Oven 40°C
UV: 210nm
Sample: Clrotimazol reference solution containing Imp-A,B &

Chromatogram showing two distinct peaks (Imp-E and Cefixime)
Excellent separation and baseline resolution

NovaSil®

01) Levocetirizine HCl IP Assay
Column: Novosil® C18 5µm, 250X4.6mm, P/N:NOVI18502S046-0
Mobile Phase: 60%V/V 0.05M KH2PO4 pH 6.0/ 40% V/V ACN
Flow Rate: 1.0 ml/min
Column Oven: Ambient
UV: 230nm
Sample: Levocetirizine HCl

Chromatogram showing sharp peak at retention time
Excellent peak shape and resolution

02) Azithromycin IP Assay
Column: NovaSil C18 (2) 5µm, 250X4.6mm, P/N:NO185025046-2
Mobile Phase: 10 volume of 3.484%W/V solution of K2HPO4 pH 6.5/ 35 volume of ACN/ 55 volume of water
Flow Rate: 1.0 ml/min
Column Oven: 70°C
UV: 215nm
Sample: Azithromycin Standard

Chromatogram showing sharp peak at retention time
Excellent peak shape and resolution

03) Amoxycillin & Potassium Clavulanate Tablet IP Assay
Column: NovaSil C18 5µm, 250X4.0mm, P/N:NO185025040-0
Mobile Phase: 5% V/V of Methanol/ 95% V/V of Phosphate Buffer pH4.4
Flow Rate: 2 ml/min
Column Oven: Ambient
UV: 220nm
Sample: Amoxycillin & Potassium Clavulanate Tablet

Chromatogram showing sharp Two well-resolved peaks (Clavulanate and Amoxycillin)

04) Ibuprofen and Paracetamol Tablets IP Assay
Column: NovaSil C8 5µm, 250X4.6mm, P/N:NO085025046-0
Mobile Phase: 75%V/V 0.075M Phosphate Buffer pH7.0/ 24.5% V/V of ACN/ 0.5%V/V of Methano
Flow Rate: 1.5ml/min
Column Oven: 35°C
UV: 220nm

back top